Vacuum Metrology & Semiconductor Load-Lock Safety, Marcello Pirani (1906) Gauge Physics
Pirani Vacuum Gauge Heat Loss & Gas Calibration Formula
Calculate Nitrogen-Equivalent Indicated Pirani Pressure (mbar), Gas-Conductive Filament Heat Loss (mW), Total Wheatstone Bridge Power (mW), and Gas Signal Share (%).
Venting a vacuum load-lock with Argon or Helium on an un-corrected N2-calibrated Pirani gauge blows viewport O-rings right off their clamps
Marcello Pirani’s (1906) Constant-Temperature Thermal-Conductivity Vacuum Gauge holds a fine tungsten/platinum sensor wire at T_w ~= 395 K inside a T_0 ~= 295 K tube using an automatic Wheatstone feedback bridge. Across 10⁻³ to 10 mbar (Knudsen molecular-to-transition regime), gas molecules colliding with the hot wire carry away **Conductive Power Qgas = (Ptrue / C_f) * K_0 * (T_w - T_0) (mW, atop constant Radiation + End-Support Loss Qrad mW), so the bridge reports Nitrogen-Equivalent Indicated Pressure Pind = Ptrue / C_f (mbar). Because heavy monatomic Argon (C_f ~= 1.62 in molecular flow; and even worse above 20 mbar viscous flow where Pirani output saturates flat)** conducts 38% *less* heat than Nitrogen (C_f = 1.00), an un-corrected Pirani gauge under-reads true Argon pressure—trick-locking PLC vent valves open until positive overpressure pops elastomer viewports.
Start on an N2-Calibrated (C_f = 1.00) Pirani Gauge Tube—True_Chamber_Pressure Ptrue = 0.850 mbar (B2), Gas_Correction_Factor C_f = 1.00 (B3), Wire_Temp T_w = 395 K (B4), Tube_Wall_Temp T_0 = 295 K (B5), Free-Molecular_Conductance K_0 = 0.48 mW/(mbar·K) (B6), and Parasitic_Radiation/End_Loss Qrad = 4.20 mW (B7). Ask the in-app assistant for a thermal-bridge energy-balance hint: Indicated Gauge Pressure equals 0.850 mbar (E2), Gas Heat Transfer equals 40.80 mW (E3), Total Filament Bridge Power equals 45.00 mW (E4), with 90.7% active gas signal above the radiation zero-floor (E5).
| Pirani Vacuum Gauge KPI (Tw=395 K, T0=295 K, K0=0.48 mW/mbar·K, Qrad=4.2 mW) | Formula | Verified Output |
|---|---|---|
| N2-Indicated Gauge Pressure (mbar) | True_Pressure_mbar / Gas_Cal_Factor_Cf | 0.850 mbar N2 (0.741 mbar @ 1.20 mbar Ar) |
| Gas Conductive Heat Loss (mW) | Indicated_P_mbar * K0 * (Wire_Tw - Wall_T0) | 40.80 mW (35.56 mW @ Argon) |
| Total Bridge Filament Power (mW) | Gas_Heat_Loss_mW + Radiation_End_Loss_mW | 45.00 mW (39.76 mW @ Argon) |
| Active Gas Signal Fraction (%) | (Gas_Heat_Loss_mW / Total_Power_mW) * 100 | 90.7% (<10% below 1e-3 mbar floor) |
Pirani Wheatstone Filament Heat Balance & Gas Calibration Factor Formula
Validate True_Pressure in [0.0005, 50] mbar, Gas_Correction C_f in [0.30, 3.00], Wall_Temp T_0 in [250, 350] K, Wire_Temp T_w >= T_0 + 30 K up to 550 K, Sensitivity K_0 > 0.05 mW/(mbar·K), and Radiation_Loss > 0.2 mW.
=LET(pt_m, B2, cf_g, B3, tw_k, B4, tz_k, B5, kz_s, B6, qr_m, B7, qg_m, (pt_m / cf_g) * kz_s * (tw_k - tz_k), IF(OR(pt_m < 0.0005, pt_m > 50, cf_g < 0.3, cf_g > 3, tz_k < 250, tz_k > 350, tw_k < tz_k + 30, tw_k > 550, kz_s <= 0.05, qr_m <= 0.2), "Invalid Pirani Input", ROUND((qg_m / (qg_m + qr_m)) * 100, 1)))Argon Sputter-Gas Backfill (P_true = 1.20 mbar, C_f = 1.62) Under-Reading Trap Check
Backfill with pure Argon to B2 = 1.20 mbar (C_f B3 = 1.62) without switching the controller gas table: Indicated Display E2 reads only 0.741 mbar (38% too low), proving why safety atmospheric vent interlocks ALWAYS pair a Pirani with a gas-independent Capacitance Manometer / Piezo Diaphragm switch.
=IF(ISTEXT(E5), E5, ROUND(B2 / B3, 3))Try the example
Use the original values or change them to test your understanding.