Pharmaceutical Quality Control & GMP Release, USP <1216> Uncoated Tablet Friability

USP <1216> Tablet Friability & Sample Size Formula

Calculate Harmonized USP <1216> Minimum Tablet Count (6.5 g vs 10-Tablet Rule), 100-Drop Abrasion Mass Loss (mg), Percentage Friability (%), and <=1.0% GMP Release Gate.

Counting 10 small 200 mg tablets violates harmonized USP <1216>—and losing over 1.0% edge dust inside the 283 mm drum blocks pan-coating release

Uncoated pharmaceutical tablets tumbled inside high-shear perforated coating pans or blister-pack feeders chip and dust unless mechanical edge strength passes harmonized USP <1216> / Ph. Eur. 2.9.7 / JP Tablet Friability (100 revolutions at 25 ± 1 RPM = 4 minutes inside a 283–291 mm acrylic scoop drum dropping tablets 156 mm onto the curved baffle every turn). First, the compendial Sample-Size Rule branches strictly at 650 mg unit tablet weight: for tablets <= 650 mg, weigh a whole-tablet sample as close as possible to >= 6.5 g (ROUNDUP(6,500 / Wunit,mg, 0) tablets); for large caplets > 650 mg, test exactly 10 whole tablets. After careful air-jet dedusting before (Winit g) and after 100 drops (Wfinal g), **Friability %Loss = ((Winit - Wfinal) / Winit) * 100% MUST NOT exceed 1.0%** (with zero cracked, cleaved, or capped tablets).

Verify on a 420 mg Uncoated Ibuprofen Core Lot—Nominal_Unit_Tablet_Mass = 420.0 mg (B2 -> <= 650 mg rule requires ROUNDUP(6500 / 420, 0) = 16 tablets in E2), Dedusted_16-Tablet_Initial_Mass Winit = 6.725 g (B3), and Post-100-Drop_Dedusted_Mass Wfinal = 6.682 g (B4): Abraded Dust Loss equals 43.0 mg (E3), Friability equals 0.639% <= 1.0% (E4), certifying "USP <1216> FRIABILITY PASS" (E5). In a shared GMP batch-record workbook, QC analysts can attach instrument .csv balance logs as non-executing excerpts while QA reviewers owner/editor verify E5 before signing lot disposition.

Harmonized USP <1216> Friability KPI (420 mg Core, W_init=6.725 g, W_final=6.682 vs 6.638 g)FormulaVerified Output
Compendial Minimum TabletsIF(Unit_mg <= 650, ROUNDUP(6500/Unit_mg, 0), 10)16 tablets (>= 6.5 g harmonized rule)
100-Drop Abraded Mass Loss (mg)(Initial_Weight_g - Final_Weight_g) * 100043.0 mg (87.0 mg @ over-dried granulation)
Tablet Friability Loss (%)((W_init_g - W_final_g) / W_init_g) * 1000.639% (1.294% > 1.0% limit)
USP <1216> Release DispositionFriability_Pct <= 1.0% Compendial CeilingUSP <1216> FRIABILITY PASS

Harmonized USP <1216> / Ph. Eur. 2.9.7 Sample-Count & 1.0% Release Formula

Validate Unit_Mass in [25, 3,000] mg, Initial_Weight Winit >= 5.0 g, and 0.80 * Winit <= Wfinal <= Winit right on E2 (checking ISTEXT(E2) on E3..E4 so E5 disposition text stays decoupled).

=LET(um_m, B2, wi_g, B3, wf_g, B4, fr_p, ((wi_g - wf_g) / wi_g) * 100, IF(OR(um_m < 25, um_m > 3000, wi_g < 5, wf_g > wi_g, wf_g < 0.8 * wi_g), "Invalid Friability Input", IF(fr_p <= 1, "USP <1216> FRIABILITY PASS", "REJECT (>1.0% LOSS OR CAPPING)")))

Over-Dried Low-Moisture Granulation (W_final = 6.638 g -> 1.294%) Lot Rejection Check (Set B4 = 6.638 g)

Over-dry fluid-bed granules below 1.0% LOD so edges chip inside the Roche drum (Wfinal B4 = 6.638 g): Mass Loss E3 doubles to 87.0 mg and Friability E4 climbs to 1.294% (> 1.0%), triggering "REJECT (>1.0% LOSS OR CAPPING)" in E5.

=IF(ISTEXT(E2), E2, ROUND(((B3 - B4) / B3) * 100, 3))

Try the example

Use the original values or change them to test your understanding.